What does it mean if my Binder Laboratory Equipment chamber is without function, and the standby icon is displayed?
- AAaron CummingsSep 8, 2025
The chamber is likely in standby mode. Press down the standby button until the display lights up.
What does it mean if my Binder Laboratory Equipment chamber is without function, and the standby icon is displayed?
The chamber is likely in standby mode. Press down the standby button until the display lights up.
What to do if there is an electrical hazard during live maintenance work on Binder Accessories?
To avoid electrical hazards during maintenance on your Binder Accessories, ensure the chamber remains dry, never remove the rear panel, and always disconnect the chamber from power by turning off the main switch (for ED, FD, FED size 720 models) and pulling the plug. It is essential that only licensed electricians or experts authorized by BINDER perform any maintenance work.
How to fix a Binder ED 260 fan that doesn’t turn or turns too slowly?
If you have a BF or FED model, the fan speed might be set too low; set the fan speed to 100%. If the problem persists, further assistance is needed.
What does the alarm message “!TProt” mean on my BD or BF Binder ED 260?
The safety device class 3.1 has responded. Check the settings of the temperature set-point and of the safety device class 3.1.
Why are there deviations from the indicated heating-up times in my Binder Laboratory Equipment?
The chamber might be fully loaded. Try loading the chamber with fewer items or consider allowing longer heating-up times.
What to do if the Binder ED 260 chamber has no function and the standby icon is displayed?
The chamber is likely in standby mode. Press down the standby button until the display lights up.
Requirement for qualified personnel, manual usage, general safety instructions, intended use, and residual risks.
Chamber overview, components, and controller operations including settings.
Unpacking, lifting, transportation, site requirements, and electrical connection.
Initial steps to power on, load, and operate the chamber.
Explanation of protective devices, safety controllers, and alarm handling procedures.
Guide to using timer functions and programming temperature ramps.
How to record data via USB and configure network settings.
Instructions for cleaning, decontaminating, and performing regular maintenance.
Guide to resolve common faults and procedures for returning equipment.
Guidelines for proper disposal of the chamber according to regulations.
Detailed technical data, dimensions, and specifications for various models.
Official EU and UL compliance statements and conformity certificates.
Forms for declaring absence of hazardous substances and requesting return authorization.