KBF / KBF P (E5.2) 02/2012 page 88/103
Mechanical safety device class
3.1 responds
(with option safety device class
3.3).
Limit temperature reached.
Check setting of temperature set-
point and of safety device class
3.1. If appropriate, select suitable
Too much external heat load.
Contact BINDER service.
Semi-conductor relay defective
Mechanical safety device class
3.2 responds
(with option safety device class
3.3).
Limit temperature reached.
Check setting of temperature set-
point and of safety device class
3.2. If appropriate, select suitable
Controller or safety device defec-
tive.
Contact BINDER service.
Refrigerating performance
Low or no refrigerating perfor-
mance.
Ambient temperature > 25 °C / 77
°F (chap.3.4).
Select cooler place of installation.
ture/humidity values not in the
optimum range (see temperature
humidity diagram, Figure 21).
Select combination of tempera-
ture/humidity values in the opti-
mum range (chap. 12).
Compressor not turned on.
Contact BINDER service.
Electro-valves defective.
No or not enough refrigerant.
Too much external heat load.
Control accuracy of ± 3 % r.H. is
Door opened very frequently. Open doors less frequently.
Humidity fluctuation, together with
temperature fluctuation
> 1 °C with a set-point approx.
3 °C above ambient temperature.
Place of installation too hot.
Select cooler place of installation
or contact BINDER service.
Low or no dehumidification.
Contact BINDER service.
Icing at the evaporator plates.
Set-point was too long below
ambient temperature.
Defrost the unit (chap. 13).
Condensation at the walls of the
inner chamber.
ture/humidity values not in the
optimum range (see temperature
humidity diagram, Figure 21)
Select combination of tempera-
ture/humidity values in the opti-
mum range (chap. 12).
Set-point was too long below
ambient temperature, icing in the
Defrost the unit (chap. 13)
Low humidity and temperature
accuracy.
Fan speed has been reduced. Set fan speed to 100%.
Buzzer at the lower part of the
unit sounds.
Water level in the steam module
is too low (filling time-out)
Turn off and on the humidity
switch (4). In case of alarm re-
peat after approx. 2 minutes,