Issue 06/2020 Art. No. 7001-0397
Operating Manual
Translation of the original operating manual
Vacuum pump VAP 1 / VAP 2
For vacuum drying oven VD
Model Voltage Equipment Art. no.
Vacuum pump VAP 1 230 V 2 heads 5013-0217
Vacuum pump VAP 1 120 V 2 heads 5013-0223
Vacuum pump VAP 2 230 V 4 heads 5013-0218
Vacuum pump VAP 2 120 V 4 heads 5013-0219
BINDER GmbH
Address: Post office box 102, 78502 Tuttlingen, Germany Phone: +49 7462 2005 0
Fax: +49 7462 2005 100 Internet: http://www.binder-world.com
E-mail: info@binder-world.com Service Hotline: +49 7462 2005 555
Service Fax: +49 7462 2005 93 555 Service E-Mail: service@binder-world.com
Service Hotline USA: +1 866 885 9794 or +1 631 224 4340 x3
Service Hotline Asia Pacific: +852 390 705 04 or +852 390 705 03
Service Hotline Russia and CIS: +7 495 988 15 16

Table of Contents

Questions and Answers

  • J
    Juan BarnesAug 12, 2025
    What to do if my Binder Water Pump does not start?
    • J
      Jeremy HolderAug 12, 2025
      If your Binder Water Pump isn't starting, there are several things to check. First, make sure the device is turned on using the main power switch. Verify the connection to the power supply. Ensure that the correct voltage is present at the socket. If the thermal switch has responded, turn off the device at the main power switch, let it cool down, and then turn it on again. If the issue persists, it could be a defective motor or a broken power cable.
  • P
    pettymichelleAug 19, 2025
    Why is my Binder VAP 1 generating no or insufficient vacuum?
    • T
      thomascamachoAug 19, 2025
      If your Binder Water Pump is not generating enough vacuum, the issue could stem from several causes. Start by checking for leaks in the connected equipment or connection elements, and seal any leaks you find, replacing seals or hoses if needed. Inspect the vacuum pump itself for leaks, particularly the hose connections between the pump heads, and replace any faulty hoses or screw connections. Also, the vacuum pump could be soiled, so general maintenance and cleaning may be required. Finally, clean the valves from condensates and impurities.
  • P
    Patrick PopeAug 27, 2025
    What to do if Binder VAP 1 valves are soiled?
    • D
      Darryl KingAug 27, 2025
      If the valves of your Binder Water Pump are soiled, clean valves from condensates and impurities.
  • J
    jason44Sep 3, 2025
    What to do if Binder VAP 1 Water Pump is leaky?
    • C
      Chad YangSep 3, 2025
      If your Binder Water Pump is leaky, check the hose connections between the pump heads and replace the hoses and / or screw connections if necessary.