VAP 1 / VAP 2 06/2020 Page 15/52
1.12 Measures to prevent accidents
The manufacturer took the following measures to prevent dangers:
• Indications on the type plate
See operating manual chap. 1.6.
• Operating manual
An operating manual is available for each device.
• Overtemperature monitoring
The device is equipped with a self-retaining thermal switch on the motor.
• Electrostatic charge
The interior parts are grounded.
The device is equipped with an IEC connector plug with a grounded plug. An electrical connection with
a protective conductor (operating mode S1: suitability for continuous operation) must be provided
(chap. 4.2).
• Non-ionizing radiation
Non-ionizing radiation is not intentionally produced, but released only for technical reasons by
electrical equipment (e.g. electric motors, power cables, solenoids). If persons with active implants
(e.g. pacemakers, defibrillators) keep a safe distance (distance of field source to implant) of 30 cm, an
influence of these implants can be excluded with high probability.
• Protection against touchable surfaces
Tested according to EN ISO 13732-1:2008.
• Floors
See operating manual chap. 3.6 for correct installation
• Cleaning
See operating manual chap. 6.2.