VAP 1 / VAP 2 06/2020 Page 22/52
Connect the vacuum hose to the suction
connection of the pump:
Fix the hose (e) on the suction connection (f) of
the pump with a hose clamp (d).
Put the second hose clamp (d) over the free
hose end.
Push the small flange (c) with the hose nozzle
(f) into the hose opening.
Push the external centering ring (b) over the
small flange (c).
Use the clamping ring (a) to connect the hose
(e) to the VD vacuum drying oven.
Figure 8: Connecting the vacuum hose
Please observe the following instructions:
Only operate the pump with the specified hose dimensions!
Make sure that the maximum permissible pressure of 1 bar at the suction connection is NOT
exceeded.
External mechanical tensions and vibrations must not be transferred to the pump.
Use a flexible hose suitable for vacuum applications to connect the pump.
Make sure that the overpressure generated at the pressure port does NOT exceed 1 bar.
NOTICE
Danger of damage due to exceeding the maximum permitted pressure.
Damage to the device.
Make sure that the maximum permissible pressure of 1 bar at the suction connection is
NOT exceeded.
Make sure that the overpressure generated at the pressure port does NOT exceed 1
bar.

Table of Contents