S3 Cell Sorter Instruction Manual98 |
Note: The user must be present to change the containers approximately 1 hour into the
procedure. After this point the system can be left overnight to complete the rinse and shutdown
if that option is chosen.
The system will soak and go through an extended warmup (Figure 120) before a system startup
is commenced (Figure 121).
After decontamination the
system should be ready
for running samples. Any
high background should be
eliminated.
Fig. 119. Decontamination window when priming rinse filter and nozzle path.
Fig. 121. Startup process at end of decontamination.
Fig. 120. Decontamination window while soaking system in proper fluidics and warming up.