Biochrom Anthos Fluido-2 Microplate Washer v2.0 Page 14
4.7 Unpacking Procedure
1. Visually inspect, the container for damage, before opening it.
Note: Report any damage immediately to the forwarding agent or to the delivery carrier.
2. Place the carton in an upright position and open it.
3. Remove the upper cardboard box carefully.
4. Lift the instrument out of the carton and place it on a flat surface, free
from dust, vibration and away from direct sunlight.
5. Visually inspect the instrument for loose, bent or broken parts.
Note: Report any damage immediately.
6. Compare the instrument's serial number, attached on the rear panel of
the instrument, against the serial number of the instrument on the
delivery (shipping) note and packing list.
6. Check the instrument accessories against the delivery (shipping) note
and packing list.
7. Please save all packing materials, as they maybe required for later
transportation.
4.8 Installation requirements, Environmental operating conditions
Flat, dry, clean and vibration-proof working area (50 cm x 35 cm -
length x width)
Additional room for cables, connections, etc.
No direct sunlight
Clean ambient air (free of corrosive vapours, smoke and dust)
Ambient temperature between 15°C and 40°C
Humidity between 15% and 85%.
Before the instrument is installed and switched on, it should be left to
stand for at least 2 hours, so there is no possibility of condensation
causing a damage or malfunction.
Use the power cable supplied with the instrument to connect the
instrument's main plug with a power outlet providing protective earth.
Please Note::
Only use the power cable supplied with the instrument or a power cable with protective
earth connection carrying the CE-mark.
The power outlet the instrument is connected to has to meet the applicable technical
safety requirements!