Biochrom Anthos Fluido-2 Microplate Washer v2.0 Page 73
11 Maintenance
11.1 Recommended regular inspection
The Fluido 2 is designed in a way that during appropriate use no danger
for the user is caused.
Correct functional performance is assured by an automatic check of the
relevant technical parameters during operation. In case of a problem the
instrument reports an error message.
Anthos recommends to have the instrument inspected by an authorized
service partner at an interval of about 2 years. During this inspection the
correct function of the instrument can be verified.
11.2 Approved parts
Except for the parts shown in the following list only parts supplied by
Anthos or an authorized Anthos Distributor may be installed in or used
with Anthos Fluido 2:
Fuses: as specified in chapter
4.1.1 Rated operating conditions
Power cable: CE-marked power cable with connected protective earth and
protective earth connector
(see also 4.8 Installation requirements, Environmental operating
conditions)
External Computer controlling the instrument
(see also Error! Reference source not found. Error! Reference source
not found.)
11.3 Exchanging a Fuse
In case of malfunction (the green stand-by indicator is not illuminated),
the fuses located in the mains inlet next to the mains socket on the rear
side of the device can be checked and replaced if necessary.
Disconnect the instrument from mains by unplugging the power cable.
Open fuse-carrier next to the mains socket with a screwdriver
Remove fuse
Insert spare fuse included in supply (2 pcs. 2.5 A TH250VAC