Biochrom Anthos Fluido-2 Microplate Washer v2.0 Page 36
Screen
15.
Waste Check
A warning to 0check the liquid level of the waste bottle is displayed (this
function needs to be activated in the "Setup Menu").
Function keys:
<F1> or <F2>
Select "Yes" or "No". If "No" is selected the instrument will return to
the main menu.
Warning: If "Yes" is selected without a sufficient volume for waste liquid in the waste
bottle the instrument can be damaged!
Screen
16.
Plate Check
A warning to check whether a plate is inserted is displayed (this function
needs to be activated in the "Setup Menu").
Function keys:
<F1> or <F2>
Select "Yes" or "No". If "No" is selected the instrument will return to
the main menu.
Warning: If "Yes" is selected without a plate inserted to the instrument spillage will be
caused which can damage the instrument and/or be harmful to the ambient
(depending on the used wash liquid)!
Screen
17.
Nr of stripes check
The display prompts you to enter the right numbers of stripes (this
function needs to be activated in the "Setup Menu").
Function keys:
<F1>
Select Esc: the instrument will return to the main menu
<->/<+>
Select number of stripes
Plate inserted?
No Yes
Nr of stripes:
Esc 12