EasyManua.ls Logo

biochrom Anthos Fluido 2 - Plate Selection; Ask for Last Strip

biochrom Anthos Fluido 2
76 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Biochrom Anthos Fluido-2 Microplate Washer v2.0 Page 46
<+> or <->
Use these keys to scroll through the available modes (“Plate” or
“Strip”).
<enter>
Confirm the selection of the processing mode and go to the next
screen.
Screen
39.
Plate Selection
Select the plate to be used for this procedure. The plate must be defined in
the "Define Plate Menu" prior to the definition of the procedure.
Function keys:
<F1> or <F2>
Select "Esc" to return to the main menu without saving any changes or
"<<" to return to the previous screen.
<+> or <->
Use these keys to scroll through the available plates.
<enter>
Confirm the selection of the plate and go to the next screen.
Screen
40.
Ask for Last Strip
Select "Yes" if the selection of a "number of strips" should be possible
every time this procedure is started. This can be useful if for example full
plates and plates with just a few strips are used subsequently with the
same procedure.
Function keys:
<F1> or <F2>
Select "Esc" to return to the main menu without saving any changes or
"<<" to return to the previous screen.
<+> or <->
Use these keys to toggle between "Yes" and "No". If "Yes" is selected
you can define the last strip every time the procedure is started. "No"
means that every time all strips of a plate are washed.
Plate:
<< Esc FB1
Ask strip nr:
<< Esc No