Biochrom Anthos Fluido-2 Microplate Washer v2.0 Page 63
Select “Mov” to move the plate transport to the current dispense
position.
<+> or <->
Use these keys to enter the dispense position. The plate can be moved
in steps of 0.1mm (range: -5.0 5.0mm).
<enter>
Confirm the dispense position and go to the next screen.
Screen
71.
Bottom Dispensing Position
For additional bottom washing the needles are moved upwards to Bottom
dispensing height and then down again to the aspiration height.
The bottom dispensing height defines the vertical position of the manifold
during dispensing of a bottom wash cycle. The dispensing height must be
higher or equal the aspiration height. It may be defined by "teach-in"
(move the manifold to the correct position with the <+> and <-> keys).
Function keys:
<F1> or <F2>
Select "Esc" to return to the main menu without saving any changes.
Select “Mov” to position the manifold to the current bottom dispense
height and switch to the “teach-in” mode.
<+> or <->
Use these keys to enter the dispense height in steps of ~0.1mm (range:
0 18.8mm).
<enter>
Confirm the bottom dispensing position and go to the next screen.
Screen
72.
Bottom Dispense Height “Teach-in” mode
In this "teach-in" mode you can move the manifold to the correct height
with the <+> and <-> keys.
Function keys:
<F1> or <F2>
Select "" to automatically detect the bottom of the well.
Bottom disp ht:
Esc Mov 0.0mm
Bottom disp ht:
0.1 mm