BM3 www.BIO2NET.com
Revision A
- 29 -
Inspection standards
- Detect the paper
- Normal operation in the speed 25mm/s, 50mm/s
- Noise when you use print
- Clear output
Inspection procedure
- Check the detection of the paper when the paper is in the device or not
- Check the speed of the printing after changing the printing speed
- Check the noise when print is working
- Check the status of the print-output
ECG
Inspection contents
- Lead Select
- Lead Fault
- Gain 1mV
- Pace Detect
- Noise
Inspection standards
- The alteration of the setting values of the lead(3 lead: I,II,III)
- Detect the each lead fault
- 10hz sine wave Amp : 10mm +/- 1mm
- Detect the pace maker signal
- 1mm
Inspection procedure