BM3 www.BIO2NET.com
Revision A
- 35 -
3.Temp malfunction
4. NIBP malfuction
Air leaks from the hose
connector or cuff
YES
NO
Replace the hose or cuff
Repair NIBP B/D
The probe is
in bad conditions
YES
NO
Replace the probe
Repair the Temp B/D
5. Abnormality in NIBP measurements
The measurements
appear lower
YES
NO
Loosen the cuff during diagnosis
There is a delay
in measuring
Repair the NIBP B/D
NO
YES
Tighten the cuff during diagnosis
The patent moved
while measuring
YES
NO
Make sure the patient stay still