BM3 www.BIO2NET.com
Revision A
- 32 -
- Check the message when the temp connector is disconnected from the
main device
NIBP
Inspection contents
- Precision
- Set the Zero
- Operate the OVP
- Auto measurement
Inspection standards
- The mean values after measuring 5 times : +/- 5mmHg
- Set the normal zero
- adult : 290mmHg
- Auto measurement in the auto-mode
Inspection procedure
- Check the values 5 times in the arm cuff mode by using the simulator ;
120/93/80, 200/196/180
- Set the zero in the device
- Check the OVP operation by using the simulator ; 290mmHg
- Check the operation in the interval 1 minute
ALARM
Inspection contents
- Control the volume of the sound
- Operate the alarm