- Precision
- No finger
- Lead Fault
Inspection standards
- Amp. 0.3 , %spo2 96: +/- 2
- Detect the NO FINGER
- Detect the lead fault
Inspection procedure
- Check the error by using the simulator ; Amp 0.3, % 96
- Check the message when you don't put the finger in the Probe
- Check the message when the spo2 connector is disconnected from the
main device
TEMP
Inspection contents
- Precision
- Lead Fault
Inspection standards
- The temp. 20, 37°C : +/- 0.1°C
- Detect the lead fault
Inspection procedure
- Check the error after setting temp. by using the precision -resistance
box