- Check the alteration after connecting the 5 leads ecg cable to
simulator
- Check the lead fault in the each lead
- Check the gain of the printer-output after setting the 10hz sine wave in
the simulator
- Check the letter P in the display after setting the pace mode in the
simulator
- Check the size of the base noise after connecting the human body
RESPERATION
Inspection contents
- Gain 1Ω
- Noise
Inspection standards
- The wave size in the LCD display and printer should be within 10mm
+/- 1 mm.
- The wave size of the noise in the LCD display and printer should be
within 2 mm.
Inspection procedure
- Check the gain of the printer-out by using the simulator ; 1Ω
- Check the noise by using the simulator ; 0Ω, display x10
SpO2
Inspection contents