BM3 www.BIO2NET.com
Revision A
- 8 -
1. Be sure that AC power supply line is appropriate to use.
(AC 100 - 240V)
2.Be sure that the power source is the one supplied from Bionet.
(DC 18V, 2.5A)
3.Be sure that the entire connection cable of the system is properly and firmly
fixed.
4. Be sure that the equipment is completely grounded.(Otherwise, noise
could result.)
5. The equipment should not be placed in the vicnity of electric generator, X-
ray, broadcasting apparatus to eliminate the electric noise during operation.
Otherwise, it may cause incorrect result.
NOTE
The Equipment should be placed far from generator, X-ray equipment, broadcasting
equipment, or transmitting wires, so as to prevent the electrical noises from being
generated during the operation, When these devices are near the Equipment, it can
produce inaccurate measurements. For BM3, both independent circuit and stable
grounding are essentially required. In the event that same power source is shared
with other electronic equipment, it can also produce inaccurate output
Warning
Do not use the System for any monitoring procedure on a patient if the monitor is not
working properly, or of it is mechanically damaged. Contact the hospital biomedical
engineer, or your supplier