22.1 Electromagnetic emissions
The BT-740 is intended for use in the electromagnetic environment specified below.
The customer or the user of the BT-740 should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The BT-740 uses RF energy only for its internal
function. Therefore, its RF emissions are very
low and are not likely to cause any interference
in nearby electronic equipment.
RF emissions
CISPR 11
Class A
The BT-740 is suitable for use in all
establishments other than domestic, and may
be used in domestic establishments and those
directly connected to the public low-voltage
power supply network that supplies buildings
used for domestic purposes, provided the
following warning is heeded:
Warning: This BT-740 is intended for use by
healthcare professionals only. This equipment/
system may cause radio interference or may
disrupt the operation of nearby equipment. It
may be necessary to take mitigation measures,
such as re-orienting or relocating the BT-740 or
shielding the location.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
Complies