44
Configure Timed Input
The load has a built-in timer function that can be configured to allow enabling (ON) the main
input for a specified amount of time. To configure this time, follow the steps below:
1. Go to CONFIG menu and select Protect. Then press .
2. Select Time and press .
3. To enable timed input, select On and press . The default is Off.
4. It will then prompt to enter a value for Delay. This is the amount of time to enable the
input from when the is pressed (or when input is turned ON remotely). Once
this time is past, input will be disabled (OFF). Use the numeric keypad or rotary knob to
enter the value. Press to confirm the change. The valid range is 0 – 60000
seconds.
5. Now, press twice to exit the menu. The instrument will show the delay timer as
follows:
Measurement Configurations
Voltage Auto Range
By default, the load has auto range enabled for voltage measurements and operations. To
enable or disable, follow the steps below:
1. Go to CONFIG menu and select Measure and press .
2. Select V-Range and press .
0.00 W 60000.0S CC = 1.000A