EasyManuals Logo

BK Precision 9115 Programming Manual

BK Precision 9115
23 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #16 background imageLoading...
Page #16 background image
[SOURce:]FALL[:LEVel]
This command is used to set the voltage falling time of the power supply.
Command syntax: [SOURce:] FALL [:LEVel] <NRf>
Unit: s
parameter: 0~65.535
Query syntax: [SOURce:] FALL [:LEVel]?
Return parameter: 0~65.535
[SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude]
This command is used to set the output current value.
Command syntax: [SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude] <NRf>
Parameter: MIN TO MAX|MIN|MAX|DEF
Unit: A |mA |uA
Query syntax: [SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude]?
Parameter: None
Return parameter: <NRf>
[SOURce:]CURRent[:LEVel]:TRIGgered[:AMPLitude]{<current level> | MINimum | MAXimum }
This command is used to set a current value to be triggered. This value will be set after the
instrument receives a trigger signal. Sending CURRent command will not impact this
command’s current setting value. Sending a query command will return the original setting
value.
Command syntax: [SOURce:]CURRent[:LEVel]:TRIGgered[:AMPLitude] <NRf>
Parameter: MIN TO MAX|MIN|MAX|DEF
Unit: A
Query syntax: [SOURce:]CURRent[:LEVel]:TRIGgered[:AMPLitude]?
Return parameter: <NRf>
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]
This command is used to set the output voltage value.
Command syntax: [SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude] <NRf>
Parameters: MIN TO MAX|MIN|MAX|DEF
Unit: V| mV| uV
Query syntax: [SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]?
Parameter: None
Return parameter: <NRf>
[SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude]{<voltage level> |MINimum | MAXimum}
This command is used to set a voltage value to be triggered. This value will be set when the
instrument receives a trigger signal. Sending VOLTage command will not impact this command’s
voltage setting value. Sending a query command will return the original setting value.
Command syntax: [SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude] <NRf>
Parameters: MIN TO MAX|MIN|MAX|DEF

Other manuals for BK Precision 9115

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the BK Precision 9115 and is the answer not in the manual?

BK Precision 9115 Specifications

General IconGeneral
BrandBK Precision
Model9115
CategoryPower Supply
LanguageEnglish

Related product manuals