2 Remote Commands
2.1 IEEE488.2 Common Commands
Here’s a list and description of all common SCPI commands supported by the instrument.
*CLS
This command clears the following registers.
Standard event register
Query event register
Operation event register
Status byte register
Error code
Command syntax: *CLS
Parameter: None
*ESE
This command can set the parameter of standard event enable register. Setting parameter can
determine which bit of standard event register is 1 and the byte will enable ESB of status byte
register as 1.
Command syntax: *ESE <NR1>
Parameter: 0~255
The value when power on: Refer to *PSC command
Example: *ESE 128
Query syntax: *ESE?
Returned parameter: <NR1>
The bit definition of the standard event enabled register:
Bit
Position
7
6
4
2
1
Bit Name
PON
Not
used
EXE
QYE
Not
used
Bit Weight
128
16
4
PON Power-on
CME Command error
EXE Execution error
DDE Device-dependent error
QYE Query error
OPC Operation complete

Other manuals for BK Precision 9115

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the BK Precision 9115 and is the answer not in the manual?

BK Precision 9115 Specifications

General IconGeneral
BrandBK Precision
Model9115
CategoryPower Supply
LanguageEnglish

Related product manuals