EasyManuals Logo

BK Precision 9115 Programming Manual

BK Precision 9115
23 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #18 background imageLoading...
Page #18 background image
This command is used to set the lower limitation of the output voltage.
Command syntax: [SOURce:]VOLTage:LIMit[:LEVel] <NRf>
Parameter: MIN TO MAX|MIN|MAX|DEF
Unit: V| mV| uV
Query syntax: [SOURce:]VOLTage:LIMit[:LEVel]?
Return parameter: <NRf>
[SOURce:]VOLTage:RANGe<voltage>
This command is used to set the upper limitation of the output voltage.
Command syntax: [SOURce:]VOLTage:RANGe <NRf>
Parameter: MIN TO MAX|MIN|MAX|DEF
Unit: V |mV |uV
Query syntax: [SOURce:]VOLTage:RANGe?
Return parameter: <NRf>
[SOURce:]APPLy {<voltage>|MIN|MAX} [,{<current> |MIN|MAX}]
This command combines two commands in one: VOLTage, CURRent. As long as the setting
value is within the range of max current and max voltage, then the output voltage and current
will execute according to the present setting value. APPLy command will only be effective when
the setting values are within the range of the upper limits. If not, an execution error will occur.
MIN and MAX can also be used as parameters. Min will enable the output voltage and current
to be 0. Max will enable the output voltage and current to be the upper limit value.
Command syntax: [SOURce:]APPLy <NRf>,<NRf>
Parameter: MIN~MAX
Unit: V, A
Query syntax: [SOURce:]APPLy?
Return parameter: <NRf>,<NRf>
2.6 MEASUREMENT Commands
MEASure[:SCALar]:VOLTage[:DC]?
This command is used to query the actual output voltage.
Command syntax: MEASure[:SCALar]:VOLTage[:DC]?
Parameter: None
Return parameter: <NRf>
Return parameter unit: V
Example: MEAS:VOLT?
FETCh:VOLTage?
This command is used to read the voltage to be handled in the sample cache. After sending the
command, the readings will be sent to the computer. This command does not affect the
instrument settings. This command does not trigger a measurement operation, and queries
only the latest available reading. Before reading the new reading, the command returns old

Other manuals for BK Precision 9115

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the BK Precision 9115 and is the answer not in the manual?

BK Precision 9115 Specifications

General IconGeneral
BrandBK Precision
Model9115
CategoryPower Supply
LanguageEnglish

Related product manuals