readings.
Command syntax: FETCh:VOLTage?
Return parameter: <NRf>
Return parameter unit: V
MEASure[:SCALar]:CURRent[:DC]?
This command is used to read the actual current.
Command syntax: MEASure[:SCALar]:CURRent[:DC]?
Parameter: None
Return parameter: <NRf>
Return parameter unit: A
example: MEAS:CURR?
FETCh:CURRent?
This command is used to read the current which is in the sample cache. After sending the
command, the readings will be sent to the computer. This command does not affect the
instrument settings. This command does not trigger a measurement operation, and queries
only the latest available reading. Before reading the new reading, the command returns the old
readings.
Command syntax: FETCh:CURRent?
Return parameter: <NRf>
MEASure[:SCALar]:POWer[:DC]?
This command is used to enquire the actual output power.
Command syntax: MEASure[:SCALar]:POWer[:DC]?
Parameter: None
Return parameter: <NRf>
Return parameter unit: W
Example: MEAS:POWer?
FETCh:POWer?
This command is used to read the power which is in the sample cache. After sending the
command, the readings will be sent to the computer. This command does not affect the
instrument settings. This command does not trigger a measurement operation, and queries
only the latest available reading. Before reading the new reading, the command returns the old
readings.
Command syntax: FETCh:POWer?
Return parameter: <NRf>

Other manuals for BK Precision 9115

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the BK Precision 9115 and is the answer not in the manual?

BK Precision 9115 Specifications

General IconGeneral
BrandBK Precision
Model9115
CategoryPower Supply
LanguageEnglish

Related product manuals