63
4. Select <Von Point> and press Enter.
on Point
Latch
5. Set the Von Point. Input 5 and press Enter to confirm.
5V
6. Press to select <Von Latch> and press Enter. Set on/off state.
Select On and press Enter to confirm.
On
Off
7.
Press Esc key twice to exit menus.
8. Press On/Off key to turn on the electronic load’s input. The
electronic load will begin sinking current at 5 V.
When in remote control mode, you can send SCPI command “VOLT:ON <n>” to set Von value; send
“VOLT:LATch ON” to enable Von LATCH function.
Save and Recall Operation
The stored settings of all channels can be recalled. The stored parameters include operation mode,
voltage/current values, slew rate, transient setting, and more. Up to 101 groups of setting
parameters can be saved. Group 0 can be used for power-up settings. Groups 1 through 100 can be
used for automatic testing parameters. All parameters are saved into non-volatile memory and will
remain saved even in a powered down state.
To Save
Panel Operations VFD Display
1. After setting up your parameters, press the Save key to save. It
will ask for the Save Group number.
0
2. Input the Save Group number (0-100) and press Enter to confirm.
1