28
3. If you need to adjust your aim, do the following:
• Press the X button. This clears the inaccurate scan results.
Note: If your instrument does not have an X button, press the DONE button and then press the SCAN
button. This resets the scan.
DONE SCAN X
READY TO SCAN
CURRENT = >132ml
• Move or angle the scanhead toward the direction of the bladder image on the Aiming icon. For example,
if the bladder image is located on the right side of the icon, aim the scanhead so the ultrasound will be
projected further to the right.
• Repeat Step 11 through Step 12 of the Measure Bladder Volume procedure. Verify the accuracy of the
scan according to the instructions in this procedure.
4. When you have collected sufcient scans in order to conrm aiming accuracy, press the DONE button.
DONE SCAN X
READY TO SCAN
CURRENT = 132ml
The Results screen is displayed. Continue to the next procedure, Conrm the Scan Results.
PRINT NOTESSCAN
21 cm
2
21 cm
2
MAIN #
132
ml