28
P063GUI001 - User guide eCelsius Performance Connect.1 - EN
Life time: 2 years (or around 500 recharge cycles).
Means to disconnect from the main supply: Unplug the power cable
ENVIRONMENTAL CONDITIONS
In operation
- Humidity between 30 and 80% of relative humidity
- Atmospheric pressure between 800hPa and 1060hPa
- Ambiant temperature between 5 and 40°C
Storage and/or transportation
- Humidity between 30 and 80% of relative humidity
- Atmospheric pressure between 700hPa and 1060hPa
- Ambiant temperature between 5 and 35°C
- Avoid sunlight exposure
- Avoid sprayed water
Storage/Transport Use
Temp °C Humidity % Pressure HPa Temp °C Humidity % Pressure HPa
eCP capsules -
P022
5 - 35°C 30% - 80%
700 -
1060hPa
25 - 45°C 100 %
700 -
1060hPa
ePerf Connect
device- P110
5 - 35°C 30% - 80%
700 -
1060hPa
5 - 40°C 15% - 90%
700 -
1060hPa
Activator -
P030
0 - 45°C 30% - 80%
700 -
1060hPa
5 - 40°C 30% - 80%
700 -
1060hPa
System
P063
5 - 35°C 30% - 80%
700 -
1060hPa
5 - 40°C 30% - 80%
700 -
1060hPa