46 Operating Your Ultrasonic Bath
Non-Cleaning Application
This method can be used for sample preparation, which includes
degassing liquids, mixing, homogenization, dissolving solids, cell lysing
and dispersion of particles.
How it works:
• Fill the tank with water and wetting agent. The tray or beakers can be
filled with any amount of solution as long as the water in the tank
outside the tray or beakers reaches the water level line.
• Place the beakers in a beaker positioning cover, an insert tray or an
insert basket to fit your unit, or place the beakers and flasks onto a
Branson support rack. Beakers should not touch the tank’s bottom.
INSERT TRAY
BEAKER POSITIONING COVER
BEAKERS