Why is my Braun Aesculap ELAN 4 Medical Equipment applied part not running?
- JJonathan AguilarAug 6, 2025
If your Braun Medical Equipment applied part is not running, several factors could be at play. First, ensure the applied part is set to the "On" position to allow operation. Then, verify that the motor hoses with hand control and wall adapters are correctly coupled. Check the building technology to confirm there is compressed air at the wall outlet. If using a gas cylinder, make sure it is not empty and the valve is open. If all these are in order, the applied part or the motor hose with manual control might be defective and need repair by the manufacturer.