11
MENU STRUCTURE / NAVIGATION
nb has been pressed while the pump is
infusing. Start manual bolus at
1200 ml/h by pressing
k (see top of
display) or proceed to set bolus limit
with l (see bottom of display).
All status information is available in the
bottom line of the display. The desired
information can be selected by using d
u
and will be displayed permanently
thereafter (e. g. drug long name, current
system pressure etc.).
Display Meaning
Pre-alarms are indicated by a message on
the display (e.g. “VTBI near end”), an
audible tone andthe yellow LED is con-
stantly on.
To confirm a pre-alarm press k.
This hint pops up if a user tries to edit or
change a parameter by pressing l when
that parameter is unable to be changed.
Press and hold o for 3 sec to turn pump
off. A white bar stretches from left to
right and counts down the 3 sec.
As long there is an infusion line inserted
the pump will not turn off but will use
standby.
Set pressure level with l or r and
confirm by pressing k.
Cancel to edit pressure by using c.
In case of an operating alarm (e.g. "VTBI
infused") the infusion stops, an audible
tone sounds and the red LED flashes.
Confirm alarm by using k.
Confirming does not activate an acoustic
feedback.