19
Chapter 1
OPERATION
Close the pump door, confirm the inserted line with kand open the roller clamp.
If required, prime the pump with u. Then press d to proceed.
Establish the patient connection and check the parameters with t.
Start the infusion by pressing sf.
Note: A new therapy can be started at any time during a stopped infusion. If the pump
is in the Main, Status or Options Menu, press
C (repeatedly) and follow the
instructions as described.
1.5 End of Infusion
Press sf to stop the infusion. The green LED goes out. Close the roller clamp
and interrupt the patient connection.
Press x. Answer the question whether the pump door is to be opened with
u.
Press down the green opening lever completely until it locks in place.
Remove the line and close the pump door.
Press o for 3 sec to switch off the pump.
Note: The settings will be permanently saved by the switched off device.
Note: Pump cannot be powered off with IV line inserted.
Remove emergency aperture cover.
Turn crank inside aperture to open
the door.