19
Chapter 1
OPERATION
• Close the pump door, confirm the inserted line with kand open the roller clamp.
•
If required, prime the pump with u. Then press d to proceed.
•
Establish the patient connection and check the parameters with t.
•
Start the infusion by pressing sf.
Note: A new therapy can be started at any time during a stopped infusion. If the pump
is in the Main, Status or Options Menu, press
C (repeatedly) and follow the
instructions as described.
1.5 End of Infusion
• Press sf to stop the infusion. The green LED goes out. Close the roller clamp
and interrupt the patient connection.
•
Press x. Answer the question whether the pump door is to be opened with
u.
•
Press down the green opening lever completely until it locks in place.
Remove the line and close the pump door.
•
Press o for 3 sec to switch off the pump.
Note: The settings will be permanently saved by the switched off device.
Note: Pump cannot be powered off with IV line inserted.
Remove emergency aperture cover.
Turn crank inside aperture to open
the door.