56
TEC HNIC AL D ATA
Type of unit Volumetric infusion pump
Classification (acc. to IEC/EN 60601-1) defibrillator-proof; CF equipment
Protective Class II; Protective Class I in
combination with SpaceStation
Class (acc. to Directive 93/42 EEC) IIb
Moisture protection IP 22 (fluid protected for horizontal usage)
External power supply:
Rated voltage Via B. Braun SpaceStation or optional
mains adaptor (rated voltage 100 240 V
AC~, 50/60 Hz) for stand alone operation
External low voltage 11 16 V DC via Connection Lead SP
12 V or via SpaceStation
Staff call Max. 24 V / 0,5 A / 24 VA (VDE 0834)
EMC IEC/EN 60601-1-2 / 60601-2-24
Time of operation 100 % (continuous operation)
Operating conditions:
Relative humidity 30 % 90 % (without condensation)
Temperature +18 +35 °C (+64° F ... +95° F)
Atmospheric pressure 500 1060 mbar
Storage conditions:
Relative humidity 20 % 90 % (without condensation)
Temperature -2C ... +5C (- F ... +13F)
Atmospheric pressure 500 1060 mbar
Type of battery pack (rechargeable) Li-Ion
NiMH
Operating time of rechargeable battery Li-Ion
Wireless active Infusomat® at 100 ml/h
typ. 4 hours
Wireless active Infusomat® at 1200 ml/h
typ. 2.5 hours
Wireless active Infusomat® at 25 ml/h
typ. 4 hours
Wireless inactive Infusomat® at 100 ml/h
typ. 12 hours
Wireless inactive Infusomat® at 1200 ml/h
5 hours
Wireless inactive Infusomat® at 25 ml/h
15 hours
NiMH
at 100 ml/h typ. 13 hours
at 1200 ml/h typ. 5 hours
at 25 ml/h typ. 16 hours
TECHNICAL DATA
Chapter 9

Related product manuals