3.8 Program Mode
Program Mode is for infusion requiring a non-standard delivery pattern. The user de-
fines a series of intervals (max. 12 intervals) by certain parameters (rate, time, volume)
for each cycle.
The pump automatically gives each programmed period, one after the other
Example:
Program Mode should only be performed by an experienced user being familiar with
the principles of the Program Mode function and properly trained in using the present
device.
Note: The active Program Mode function always displays this icon in the Display
( ).
Note: Bolus function is disabled for Program Mode.
Set Profile Parameters: The therapy can be started directly via the drug library or via
the Main Menu/Special functions.
Starting Program Mode via Drug Library:
Note: Program Mode settings have been configured in the Drug List Manager
before and have been uploaded into the pump.
• Switch on pump with
o and wait until self-check is finished.
• Insert disposable and use the drug library according to Instructions for Use.
• Select the desired drug with
t and press l.
• Select Program Mode with
l.
In the following screen the user has to confirm the number of steps for the
therapy with
k.
SPECIAL FUNCTIONS