23
Chapter 3
SPECIAL FUNCTIONS
a Enter the concentration by entering the amount of the active ingredient
and the volume. In order to do so set the values with
q and confirm
with
k.
a If the patient's weight does not need to be entered press
l.
Press
t to choose “weight” or “surface” and confirm with k.
a Set the patient weight with
q and confirm with k.
a Select the dose prescription with
t and confirm it with l.
a Set the dose with
q and confirm with k. The rate will be
automatically calculated and displayed at the bottom of the display.
a Check the calculated rate and if necessary the adapted parameters with
t on plausibility before starting the infusion with sf.
a Check the parameters with
t on plausibility before starting with sf.
Dose can belatedly be changed in the Main Menu in the same way as the rate,
VTBI and time (compare 2.2). The effect of dose
modifications on other parameters is shown at the bottom of the display.
Additionally the total and intermediate amount of the infused drug can be
taken from the Status Menu. These can be checked and reset in the same way
as the other total and intermediate values.
A deactivation of the dose rate calculation is only possible when the pump is
stopped. Press
c from Main Menu and then press d.
Caution: A change of the patient weight or height will alter the flow rate.
3.3 Drug Library
Up to 1200 drug names including therapy data, information and up to 10 con-
centrations per drug can be stored in 30 categories. These drugs can be subdi-
vided in 50 care units and 16 Patient Profiles. The loading process into the
pump can be performed via a separate PC program (Space Upload Manager &
HiBaSeD).
Note: The drug library can be started over the Start Up and Special Functions
Menu. The user has to make sure prior to the therapy start that the drug library
in the pump complies with the patient target group. The name of the care unit
and creation date (see headline) should be checked in the pump.
There are different ways assigning a drug to an infusion. This can be done while
the infusion is running or when the pump is stopped.
On the one hand, a drug name including the according therapy data can be
taken from the drug library. On the other hand, if a rate, VTBI and/or time were
already defined in the Main Menu, the drug name and the adjusted values of
the data set will be loaded. If a dose rate calculation has already been started a
belated assignment of the drug name nevertheless is possible.