OPERATION
Chapter 1
Caution: Please make sure that the line is straight and not slack to prevent a damage
occurring to the infusion line which leads to an incorrect dosage.
Firmly press tubing into the air sensor guide to make sure the line is properly
inserted into the sensors. Thread the tubing through the notches on the right and
left side of the pump.
Close the pump door by firmly placing pressure with both hands on each side of the
pump door, continue to press firmly until you hear and feel the motorized door
latch pull the door shut. Do not open roller clamp until the pump directs you to do
so when self test is completed. Then select the inserted line with
t and confirm it
with
l. Open the roller clamp.
Caution: Before opening door, please close roller clamp and ensure door does not fall
open. If door opens to the horizontal position, please check that the slide clamp (green)
1. Mind flow direction while
inserting the infusion line
(s. pictogramm) to prevent the
danger of backflow.
2. Completely insert and then gently
press the infusion line into the
guidance and the sensors.
3. Take care that the line is perfectly
engaged with the drive and the
sensors. Especially the air sensor
needs perfect contact to the line!
Press down the opening lever.