45
Level 1:
A modification of values as well as a bolus application are not possible but a
change of the disposable can be conducted. It is possible to navigate through all
menus and status data can be checked. Starting, interrupting and switching the
pump off is possible.
Level 2:
This level has the same performance characteristic as described under level 1
and additional will not allow the change of disposable. In order to prevent a
data lock alarm the correct code must be entered within 20 sec after the pump
was stopped. Changing the disposable and switching the pump off is only
possible after the code was entered.
Level 3:
This level will allow starting and stopping the pump as well as switching off.
The code for this level may be different for each drug and is defined in the drug
list. A change of the disposable, however, is possible by using the code defined
for the other levels. An overview about the differences between the levels 1, 2
and 3 is given by the following table.
Activation of the function:
Open data lock in Options Menu with
l.
Select between level 1, 2 or 3 (if activated) with
l and r and confirm
with
k.
Enter code with
q and press k in order to activate data lock.
Changes to the protected values and the bolus function which are marked with
y are only possible after entering the code. After 20 sec in the Main Menu,
Status Menu, Special Functions Menu and Options Menu the lock will be acti-
vated again. If the wrong code is entered twice the pump will switch into the
last menu. If the wrong code is entered twice again the pump will go into an
audible alarm, a nurse call will go off and the yellow LED blinks. If a target
value was reached while data lock is active a new start of the pump is only
possible after entering the code.
Event
Change of disposable
Level 1 Level 2 Level 3
Change of parameters
O
OO
O
Stop of infusion
%
Switching off pump / Standby
O O
PCA bolus with pump-based bolus button
= possible | O = not possible | % = followed by standby-alarm
O O
Customisable screen
O O
Acoustic feedback of demanded boli
O O
Indicates denied PCA boli
O
Start of infusion
O with code for
level 1/2