REMOVING/REPLACING THE MAIN COMPONENTS PV 101+/PV 102+ Service Manual
3–2 BREAS MEDICAL Doc. No. 003259 En Issue: X-1
3.2 Removing and reassembling the upper casing
Disconnect the power supply before removing the casing of the machine.
To remove the casing:
1 Remove the mains power cord.
2 Turn the PV 101+/PV 102+ upside down. Remove the four screws placed in the
four holes underneath the lower casing and the two screws placed underneath the
handle.
3 Lift up the upper casing. Disconnect the CPU board’s ribbon cable from the MDA
board. Put the upper casing aside.
To reassemble the casing:
• Reassemble in reverse order.