EasyManua.ls Logo

Breas PV 102+ - Removing and Replacing the Power Board

Breas PV 102+
90 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PV 101+/PV 102+ Service Manual REMOVING/REPLACING THE MAIN COMPONENTS
Doc. No. 003259 En Issue: X-1 BREAS MEDICAL 3–7
3.6 Removing and replacing the power board
To remove the power board:
1 Remove the upper casing (See section 3.2).
2 Remove the main assembly (See section 3.4).
3 Remove the blower assembly (See section 3.5).
4 Move aside the patient air and sensor tubing. Disconnect the power cables to the
external DC and mains supplies on the MDA board.
5 Disconnect the DC control cable from the PSU board.
6 Unscrew the two screws that hold the rear part of the power board, and remove
the two nuts under the base plate.
7 Pinch the two securing pins on the front part of the power board, e.g. by using a
screwdriver, and lift up the power board from the base plate.
To reassemble the power board:
Reassemble in reverse order.
2 × Phillips 1
6
4
7
5

Table of Contents