EasyManua.ls Logo

Breas PV 102+ - Service Record for BREAS PV 101+;PV 102

Breas PV 102+
90 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Doc. No. 003259 En Issue: X-1 BREAS MEDICAL 9–3
PV 101+/PV 102+ Service Manual APPENDICES
Service Record for BREAS PV 101+/PV 102+
Use a photocopy of this service record for the maintenance inspection described
in chapter 2 in this service manual. Use the reverse side for comments and notes.
Service Record No:__________
Registration
Model:................................... Serial No................................... Inventory No. ..................
Accessories: ...............................................................................................................................
Delivery date:...................................... No. of operating hours:..............................................
Service started:................................... Signature: ..................................................................
Service completed:............................. Signature: ..................................................................
Delivered, date: .................................. Signature: ..................................................................
See instruction ref. Check OK
Preparing inspection
Verifying the components and the installed software 2.3.1 .........
Initial recording 2.3.2 .........
Markings 2.3.3 .........
Information from the patient/user 2.3.4 .........
Validity of the documentation 2.3.5 .........
External inspection
Visual inspection for external damage and wear 2.4.1 .........
Checking the power supply connection 2.4.2 .........
Short function check 2.4.3 .........
Internal inspection
Cleaning the inside of the machine 2.5.2 .........
Inspecting the internal wiring 2.5.3 .........
Securing the components 2.5.4 .........
Checking the power supply 2.5.5 .........
Checking Electrical safety measurements 7.5 .........
Bl
ower Assembly (15,000 hours)
Checking if blower unit should be replaced 3.5 .........
Calibration (12 months/15,000 hours)
Calibrating the pressure sensor 4.3 .........
Calibrating the flow sensor 4.5 .........
Calibrating the high pressure limit 4.6 .........
Continued on next page