Model 871A User’s Manual REMOTE COMMUNICATIONS
Bristol Instruments, Inc. 42
Connecting Using Telnet
A Telnet session example from a networked PC running Windows 10, 8, or 7 is given
below.
NOTE: The 871 will only support a single Telnet session per instrument.
1 Click on Start/Programs/Accessories/Command Prompt, or type “cmd” into the
Search bar.
2 To start a Telnet session, type “Telnet” at the command prompt followed by the 871
system’s IP address displayed in the Instrument Menu of the NuView software.
C:\>Telnet 192.168.1.109
MityDSP Telnet Server
Copyright 2012, Critical Link LLC
Ctrl-D Exit
Ctrl-E _ Toggle Echo
Bristol Instruments
*IDN?
BRISTOL WAVELENGTH METER, 871A-VIS, 1109, 1.2.11
:MEAS:POW?
0.5144
Telnet must be explicitly installed on a PC running Windows Vista, using the following
instructions.
1 Go to the Start Menu, then click Control Panel.
2 Select Programs and Features.
3 Click “Turn Windows features on or off,” found in the left column.
4 Check the box next to Telnet Client, and click OK.