Installation and Operation Manual
X-TMF-SLA5800-Series-RevB-MFC-eng
Part Number: 541B187AAG
January, 2017
Brooks
®
Digital MFC's & MFM's
Brooks
®
SLA5800 Series
Elastomer Sealed, Digital,
Thermal Mass Flow Meters and Controllers
Model SLA5850
with RS485 Communications
Model SLA5850
with DeviceNet
®
Communications
Model SLA5850
with Profibus
®
Communications
Model SLA5851
with EtherCAT
®
Communications

Table of Contents

Questions and Answers

  • N
    nreedSep 10, 2025
    What to do if Brooks SLA5863 output signal stays at approx. 5.5 Vdc or 22 mA?
    • H
      Heather StoneSep 12, 2025
      If the output signal of your Brooks Measuring Instruments stays around 5.5 Vdc or 22 mA, even when the Setpoint changes, and there's flow, it could be due to a valve issue. Specifically, the valve may be leaking or stuck open. Another potential cause is +15 V applied to the valve override input. You may also check the valve override terminal (Pin 12).
  • A
    Aaron PhillipsSep 12, 2025
    How to fix controller oscillations in Brooks SLA5863?
    • A
      Andrew Davis Jr.Sep 12, 2025
      If the controller oscillates in your Brooks Measuring Instruments, it could be due to pressure drop or inlet pressure deviating from calibrated values. Adjust the pressures to the original specifications.
  • C
    Cheryl HopkinsSep 14, 2025
    Why won't the output signal go below 2% on Brooks Measuring Instruments?
    • J
      Jennifer WareSep 14, 2025
      If the output signal of your Brooks Measuring Instruments follows the setpoint at higher setpoints but will not go below 2%, it might be because the control valve leaks or is stuck open. You should clean the control valve.
  • S
    Shelby JonesSep 16, 2025
    What causes low flow on Brooks Measuring Instruments?
    • D
      Devin RamseySep 16, 2025
      If the instrument is grossly out of calibration and the flow is lower than desired in your Brooks Measuring Instruments, it could be due to a partially clogged restrictor. Replace or clean the restrictor.

Summary

Section 1 Introduction

1-4 Specifications

References tables for detailed technical specifications of the SLA5800 Series, including performance ratings.

Section 2 Installation

2-9 Installation

Provides detailed CAUTIONS and recommended procedures for installing the MFC/MFM in a clean, dry environment.

Special considerations for high pressure installations

Discusses managing pressure during startup for high-pressure applications to prevent damage.

Special considerations to be taken when installing the SLA5853 MFC

Highlights specific installation advice for the SLA5853 model due to its dual-stage valve design.

2-11 Operation Check Procedure (Analog I;O)

Provides step-by-step instructions for verifying the proper operation of the MFC/MFM using analog I/O signals.

2-12 Operation Check Procedure (Digital I;O)

Outlines the procedure for checking the operation of MFC/MFM via digital I/O and network controllers.

Section 3 Operation

3-6 Alarms (Analog versions only)

Outlines alarm features for analog versions, including diagnostic and general alarms, and their configuration.

Section 4 Maintenance & Troubleshooting

4-1-1 Troubleshooting Analog or Digital

Offers guidance on diagnosing common MFC/MFM problems related to flow rate and signal issues.

Section A Essential Instructions

General Safety Instructions

Covers crucial safety warnings and precautions for installing and operating Brooks Instrument products.

European Pressure Equipment Directive (PED)

Explains the applicability of the PED directive to pressure equipment and compliance requirements.

European Electromagnetic Compatibility (EMC)

Details EMC directive compliance and requirements for signal cable quality and connectors.

Electrostatic Discharge (ESD)

Provides essential handling procedures for components sensitive to electrostatic discharge to prevent damage.

Brooks SLA5863 Specifications

General IconGeneral
ModelSLA5863
ManufacturerBrooks
TypeMass Flow Controller
Flow RangeDepends on model and configuration
Pressure RatingDepends on model and configuration
Operating Temperature0 to 50°C
Output Signal0-5 VDC, 4-20 mA
Power Supply15-24 VDC
MaterialsStainless steel
ApprovalsCE
Response Time< 1 second
Communication ProtocolAnalog
Flow Rate RangeDepends on model and configuration
Gas CompatibilityVarious gases (specific gases depend on configuration)