81
ZTKS0162 / Z31928 / 04
Technical Data MS 400‘54 Ascend
Fringe Field Plot
Figure A.9: Fringe Field Plot
Fringe Field Unit R max Unit Z max Unit
200 mT 0.21 m 0.34 m
5.0 mT 0.38 m 0.64 m
3.0 mT 0.41 m 0.70 m
1.0 mT 0.46 m 0.87 m
0.5 mT 0.50 m 1.00 m
0.2 mT 0.65 m 1.24 m
0.1 mT 0.83 m 1.47 m
0.05 mT 1.08 m 1.77 m
max magnetic field B0 at access point 280 mT
max field gradient dB/dz at access point 5.4 T/m
Table A.10: Fringe Field Values