EasyManuals Logo

Bruker NMR 400 89 Ascend DNP User Manual

Bruker NMR 400 89 Ascend DNP
106 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #32 background imageLoading...
Page #32 background image
32
ZTKS0243 / Z31993 / 03
Transportation
3.5 Transportation with a Crane
A crane is recommended for lifting the cryostat out of the box.
Approved Persons: Approved crane operator
Precondition: The crane must be approved for the transportation weight (see
Appendix, Technical Data).
Attachment Points
Figure 3.4: Attachment Points for Lifting Equipment
1. Exclusively use the marked eyelets as
attachment points for the lifting equipment.
2. Use all eyelets for the lifting equipment.
Figure 3.5: Instruction Label for Lifting Equipment
3. Follow the instructions on the label on top of
the cryostat. This label gives important
information about correct attachment and
transportation.
4. Check for correct fastening of the lifting
equipment before lifting the cryostat.
5. Make sure that any movement of the crane
is as slowly as possible to avoid any
damage due to acceleration.
6. Check for correct leveling of the cryostat
while hanging on the crane.