Troubleshooting
6.2 Problem
6.2.1 During Transportation
6.2.2 During Assembling
Continued on next page
Indicator Possible reason Solution By
Tilt Watch / Shock 
Watch activated.
Careless  
transportation.
1. Accept delivery with 
reservation.
2. Remark the extent of 
damage in the trans-
portation documents.
3. Start complaint process.
Approved  
Customer  
Personnel
Visible damage. Careless 
transportation.
1. Accept delivery with 
reservation.
2. Remark the extent of 
damage in the trans-
portation documents.
3. Start complaint process.
Approved  
Customer  
Personnel
Indicator Possible reason Solution By
Ceiling height too 
low for assembling 
on magnet stand.
Site does not meet 
the required 
conditions.
Choose another site that 
meets the required 
conditions.
Bruker Service
Ceiling height too 
low for inserting the 
Helium Level Sensor
Site does not meet 
the required 
conditions.
Insert the Helium Level 
Sensor before mounting the 
magnet stand.
Bruker Service
Helium bore tube 
and radiation shield 
are not concentric.
Alignment is not 
correct.
Check fixation of the align-
ment rods.
Bruker Service