MA201435_13.1 [Original Instructions] 6/25/2012
SimpliMet 3000 Recommended Parameters
Note: The listed parameters are for single molds only. Double all heat times when using a duplex mold
assembly.
Due to variable voltages, heating times may vary. Voltages below 120 VAC will take longer
than 120 VAC and 240 VAC.
Heat and Cool times is in minutes.
1"
25 mm
1.25"
30 mm
1.5"
40 mm
2"
50 mm
Materials
Temperature
Pressure
Heat/Cool
Heat/Cool
Heat/Cool
Heat/Cool
Phenolic
150° C
(300° F)
290 bar
4200 psi
1.0 / 3.0
1.0 / 3.0
1.0 / 3.0
2.0 / 3.0
Diallyl
Phthalate
150° C
(300° F)
290 bar
4200 psi
1.0 / 3.0
1.0 / 3.0
1.0 / 3.0
2.0 / 3.0
EpoMet
150° C
(300° F)
290 bar
4200 psi
1.0 / 3.0
1.0 / 3.0
1.0 / 3.0
2.0 / 3.0
KonductoMet
150° C
(300° F)
290 bar
4200 psi
1.0 / 3.0
1.0 / 3.0
1.5 / 3.0
2.0 / 3.5
ProbeMet
150° C
(300° F)
290 bar
4200 psi
1.0 / 3.0
1.0 / 3.0
1.5 / 3.0
2.0 / 3.5
TransOptic**
177° C
(350° F)
145 bar
2100 psi
2.0 / 11.0
2.0 / 11.0
2.0 / 11.0
3.0 /11.0
** TransOptic heat and cool times are based on .5-inch thick finished mount. For thicker mounts, heat times remain
the same but increase the cool time.