EasyManua.ls Logo

Burke TRI-FLEX II - CPR Procedure; Patient Restraint

Burke TRI-FLEX II
26 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
TRI-FLEX II BED - OPERATION MANUAL 11
Operational Instructions
Tri-Flex II with Fold Down Side Rails
To lower the fold down side rail, hold the top of the side and
pull up the spring loaded pull pin. Rotate the side rail away
from the bed. Allow the side rail to rotate downward until it
comes to a stop. To raise this type of side rail, lift upward on
the side rail to rotate fully upward until the spring loaded pull
pin engages the stop (See Figure 4).
WARNING: Ensure lock pin is fully engaged.
Tri-Flex II with Roll Out Side Rails
To lower the head side rail, hold the side rail at the handle and pull the spring loaded pull pin. Using the handle
on the top of the side rail, rotate the side rail toward the foot of the bed. Allow the side rail to rotate downward
untilitcomestoastop(SeeFigure3pg.12andFigure5).
To raise the head side rail, lift upwards on the handle at the top of the side rail. Allow the side rail to rotate
completely upward until the spring loaded pull pin fully engages the stop..
To lower the foot side rail, hold the top of the side rail and pull the spring loaded pull pin. Rotate the side rail
towardthefootofthebed.Allowthesiderailtorotatedownwarduntilitcomestoastop(SeeFigure5).
To raise the foot side rail lift upward on the side rail. Allow the side to completely rotate upward until the spring
loadedpin/lockengagesthestop.(SeeFigure5)
WARNING: Ensure lock pin is fully engaged.
CPR Procedures
Thisprocedurewilllowertheheadgatchinabout3-5
seconds. The emergency CPR release can be used
to lower the patient faster than running the actuator alone.
• Remove Headboard and stand at head of bed,
•PullRedCPRHandleTowardsyouapprox1-3inches.
•ContinuepullingCPRHandlewhileHeadGatchlowers.
•ReleaseCPRHandleoncebedhasloweredcompletely.
For maximum speed you should use the head down
button on the pendent control and pull the CPR release
handle on the head gatch at the same time (See Figure 6)
Figure 5
Figure 6
Figure 4
CPR Handle
!
!
!
Patient Restraints
WARNING: If patient restraints are used, they must be
attached to some point on the bed that articulates
with the
gatches when they are raised or lowered.
!