14 TRI-FLEX II BED - OPERATION MANUAL
Cleaning & Disinfection
CLEANING AND DISINFECTING
Cleaning and Disinfection are two separate procedures.
For sanitary maintenance of the bed, follow the suggested recommendations of the
manufacturer(Burke) and the institution regarding cleaner/disinfectant solutions of choice
and appropriate concentrations.
Routine cleaning is to be done on a daily basis by wiping the surfaces with a disposable
cloth using warm water and a mild detergent or cleaning solution. In the event of known infec-
tion,bloodorbodyuidscontamination,cleanrst,then disinfect following facility guidelines
using disinfectants at their recommended concentrations.
Burke recommends a chlorine based solution no greater than 10,000 PPM for routine decon-
tamination.
The bed and components should also be cleaned and disinfected between patients and be-
fore returning to service after a period of storage, per facility guidelines.
UNPLUG ALL POWER CORDS BEFORE PERFORMING CLEANING
All staff performing cleaning/disinfection processes should wear appropriate personal
protective equipment (PPE)
Do Not use alkaline or acid chemical products outside of the optimal pH range or 6-9.
Do Not use any scouring or abrasive cleaners on the bed.
Do Not use caustic or corrosive substances or chemical substances which may affect the
structure of the bed, such as acetone.
Do Not hose down, use high pressure spray or steam to clean as this may cause damage
to the electronics system and void the warranty.
Always wipe down the areas cleaned with clear water after cleaning or disinfecting
All cleaning solutions must be properly diluted according to the manufacturer’s instructions.
Follow standard institutional wipe down and infection control procedures. If the facility has
aHygieneOfcerorothergoverningbodyforcleaningandinfectioncontrolconsultthemto
ensure both internal and external compliance.
Allow all surfaces to thoroughly dry, including any electrical components that could have
become wet, before reconnecting power cord.
SEE MATTRESS/OPTION MANUALS FOR MORE SPECIFIC INDEPTH