TRI-FLEX II BED - OPERATION MANUAL 23
Assembly/ Disassembly Instruction
7. Lock the four transport casters to prevent them from allowing the bed to roll away
from you when you lift the head and foot sections.
The locked transport casters should be facing toward their respective end of the bed
(seegure15).
8. Press the Hi-Low “Down” button on the pendant until the weight of the bed rests on the
Transport Casters in the center of the bed. The Hi-Lo bed casters will lift off the ground once
the weight of the bed is on the Transport Wheels.
9. Unplug the AC power cord.
10. Release the black safety strap at the Head End of the bed and remove the Turnbuckle Tool.
11. Loosen and completely remove the two turnbuckles with the wrench. Just as in step 9 of
(Unfolding the Bed), the Turnbuckle Tool can be used as a lever to operate the turnbuckles.
12.StoretheToolintheVelcroHolderattheHeadofbedwhennishedwithtool.
13. Make sure that the black woven safety strap located at the head end of the bed is securing
the head gatch patient support surface to the base. Unexpected movement of the patient
support surface could occur if the safety strap is not securely in place.
14.Conrmthatthelocksaresetproperlyonthetransportcasters.
15.Carefullyraiseheadend,locktotransportpoleandsecureinplace
with the transport latch bracket. Take care to avoid damage of the pendant.
17. Carefully raise foot end and lock to transport pole with the transport latch bracket.
18. Secure the AC power cord to holder.
19. Store the head and foot boards on the bed as well as the Foot Side Rails.
(SeeFigure15.)
Figure 15
!