EasyManua.ls Logo

Burkert 1067 - Function of Keys in the Configuration Level; Notes on the Basic and Additional Functions

Burkert 1067
50 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1067
POSITIONER TYPE 1067
E-25-
4 OPERATION
4.5.3 Function of keys in the configuration level
Operation of the Arrow up key
- Scroll upwards in menu (selection).
- Incrementing numerical values in a selected and confirmed menu item.
Operation of the Arrow down key
- Scrolling downwards in the menu (selection).
- Decrementing numerical values in a selected and confirmed menu item.
Operation of the MANUAL/AUTOMATIC key within the main menu
- Confirmation of a selected menu item.
- Confirmation of a set value.
Operation of the MANUAL/AUTOMATIC key within the additional menu
- Confirmation of a selected menu item of the additional menu for inclusion in the main menu. The
selected menu item is marked with a star (*) in the additional menu. The menu item now appears
in the main menu where it can be selected and manipulated.
- Confirmation of a selected menu item of the additional menu, marked with a star, for deletion from
the main menu.
4.5.4 Notes on the basic and additional functions
X-SENS (factory setting: DIRECT): Specification of the type of information transfer between the
continuous action valve (valve setting) and the travel measuring system.
Options:
DIRECT: There is a linear relationship between the valve position and the input signal of the
path-measuring system.
Examples:
Fitting the positioner to a piston valve (e.g. the Type 2031) and using the external feedback/positional
transducer (linear potentiometer) to measure the piston position (see section 3.1.2 and Fig. 11).
Here, the linear movement of the piston is transformed into a linear movement of the potentiometer.
A characteristic correction is therefore not required.
Fitting the positioner to a flap valve with part-turn valve actuating drive (e.g. the type 3210) and using
the internal feedback/positional transducer (rotary potentiometer) to measure the flap position (see
section 3.1.3 and Fig. 15). The rotary movement of the flap is converted into a proportional rotary
movement of the potentiometer. A characteristic correction is not required.
LEVER: There is a sinusoidal relationship between the valve position and the input signal of
the path-measuring system.
Example:
Fitting the positioner to a membrane valve (e.g. the type 265) and using the internal feedback/
positional transducer (
rotary potentiometer
) to measure the valve position (see section 3.1.1 and
Fig. 7). Coupling is via a lever conforming to NAMUR. The linear movement of the membrane is
transformed into a rotary movement of the potentiometer. A sinusoidal curve results. In this option,
therefore, the transmission characteristic is linearized internally.
INPUT (factory setting: 4 - 20 mA): Specification of the selected unit signal.
Options:
4 - 20 mA: Use of the 4 ... 20 mA unit signal input
0 - 20 mA: Use of the 0 ... 20 mA unit signal input
0 - 10 V: Use of the 0 ... 10 V unit signal input

Other manuals for Burkert 1067

Related product manuals