12
EN
App screen Error Cause Remedy
fault
Device not working Pressure sensor for
permeate is defec-
tive or has no power
supply
Check power supply
and, if necessary,
replace the pressure
sensor or inform
service staff.
5 Repairs and maintenance
5.1 Maintenance and wearing parts
You have purchased a product that is durable
and easy to service. However, all technical
equipment requires regular servicing so that it
continues to function properly.
If the product malfunctions during the warranty
period, contact your contract partner or the
installation company, and quote the unit type
and serial number.
Observe: Please also observe the pro-
visions in the operating manual of the
installed external pre-filter.
Before performing work on electrical
components or opening the housing,
it is mandatory that the mains plug
be pulled and both the water infeed
and the permeate line be closed off in
order to guarantee that the unit is volt-
age-free.
Whenever maintenance is performed,
the connection lines and the device
must be checked for damage.
Note: Only customer service may
replace the wearing parts (e.g. after a
faulty pump).
Replacement of wearing parts:
Maintenance work: Responsible: Recommended
maintenance
interval:
General visual
inspection
Customer Monthly
Leak-tightness
inspection
Customer Monthly
Cleaning with a
damp cloth
Customer As necessary
Conductivity (with
external measuring device)
Customer/
service
At least once
per year
Replacement of the
external pre- filter
insert (particle filter
[ optionally available])
Customer/
service
Depending on
the pre-filter
used
Replacement of the
RO cartridge
Service Once per year
(recommended)
Observe: In accordance with BVG A3
(VBG4), the unit must be tested for
electrical safety every 4 years.
The BWT THERO 90PRO reverse
osmosis unit is subject to the Pres-
sure Equipment Directive 2014/68/
EU from 27 June 2014. The RO unit
fulfils the requirements set forth in arti-
cle 3, section 3, and was designed and
manu factured according to the appli-
cable engineering good practices.
The BWT THERO 90PRO device
does not have a CE mark according
to article 6, section 5, of the direc-
tive 2014/68/EU, but the attached CE
Declaration of Conformity is valid.