Action Description
Rest Time
(rest for 000:00:00)
Applies no current for the specified time period
(hhh:mm:ss)*.
The time must be specified or the cycle is skipped.
OhmTest Performs the OhmTest on the battery.
Test
The test cycle has three parts:
• The IF statement, which describes a condition that must be either true or false
for the battery being tested.
• The THEN statement, which describes the action to be taken if the battery
passes the test (the condition is true).
• The ELSE statement, which describes the action to be taken if the battery fails
the test (the condition is false).
IF Statements
Choose from ten different test conditions, or IF statements, when setting tests for
phases 1 through 5.
IF Statement Description
Skip Test Selects the TRUE statement. The FALSE statement is
ignored.
IF Target Capacity Met Selects the TRUE statement if the capacity during the
last discharge cycle meets the target capacity;
otherwise, selects the FALSE statement.
IF Target Capacity Not
Met
Selects the TRUE statement if the capacity during the
last discharge is below the target capacity;
otherwise, selects the FALSE statement.
Cadex Electronics Inc. C7000-C Series Battery Analyzers .
102